PDS_VERSION_ID = PDS3 LABEL_REVISION_NOTE = "2010-05-27, SONC, initial release " RECORD_TYPE = STREAM OBJECT = INSTRUMENT INSTRUMENT_HOST_ID = RL INSTRUMENT_ID = "PTOLEMY" OBJECT = INSTRUMENT_INFORMATION INSTRUMENT_NAME = "PTOLEMY - GAS CHROMATOGRAPH ISOTOPE RATIO MASS SPECTROMETER" INSTRUMENT_TYPE = "GAS DETECTOR" /* RLC was "GAS ISOTOPE RATIO ANALYSER" */ INSTRUMENT_DESC = " Instrument Overview =================== Ptolemy is part of the Rosetta instrument payload and is one of two evolved gas analysers of the Lander space-craft Philae. It provides data on the chemical and isotopic composition of the nucleus of comet 67P/Churyumov-Gerasimenko. Scientific Objectives ===================== The aims of Ptolemy are i) the determination of the chemical composition (accuracy 10% and mass range 10-150 Daltons) and ii) light element (H, C, N, O) isotopic composition of major components with an accuracy of 1% of volatile and refractory cometary components Calibration =========== The calibration of individual components (pressure sensors and temperature sensors) is incorporated in the EGSE software. Mass calibration of the mass spectrometer is achieved by using argon gas as a mass marker in the helium gas supply system. Operational Considerations ========================== Detectors ========= The detector is an ion trap mass spectrometer. Electronics =========== Details on the electronics of the instrument are described in the Ptolemy ADP. Location ======== Ptolemy is mounted on the large Pi-plate in the warm compartment of Philae. Samples are delivered by the Sample Drilling and Distribution unit (SD2) to ovens located on the balcony of Philae. Operational Modes ================= Ptolemy operates in three distinct modes; Safe, Standby and Science. Housekeeping data is collected during all three modes. On power on, Ptolemy enters Safe mode. Safe mode is used for processor memory management such as loading new science sequences and check memory commands. The only mode change is from Safe to Standby mode. Standby mode is used to enable components for use during the subsequent Science mode. Ptolemy can be commanded to operate a Science mode sequence or be returned to Safe mode. Science mode operates a previously loaded sequence. At the successful completion of a Science mode sequence, Ptolemy returns to Standby mode. Subsystems ========== The subsystems are: Sample Collection (SC), Gas Distribution and Processing System (GDPS), Gas Chromatography (GC), and Mass Spectrometer (MS). Sample collection consists mainly of the SD2 instrument which acquires solid samples and distributes them to Ptolemy ovens on the carousel. Gaseous samples can also be collected by an oven filled with molecular sieve or be analysed directly by the mass spectrometer. The Gas Distribution and Processing System consists of heaters, valves temperature and pressure sensors in order to prepare the sample for direct MS analysis or for injection into one of three columns of the GC system. The GDPS also contains reference materials so that the comet samples can be directly compared with terrestrial compounds of known isotopic value. The Gas Chromatography system includes 3 GC columns to separate the sample into individual components as well as the helium carrier gas required for gas chromatography. The Mass spectrometer analyses compounds giving a mass spectrum as they are eluted from a GC column. The sub-systems are described in more detail in Morse et al, 2009. Measured Parameters =================== The Ptolemy GC measures the retention times of molecular species in separation columns, the Ptolemy MS measures the mass (Daltons) of ionised fragments of molecular species. The temperature and pressure of various components are monitored. " END_OBJECT = INSTRUMENT_INFORMATION OBJECT = INSTRUMENT_REFERENCE_INFO REFERENCE_KEY_ID ="WRIGHTETAL2007" END_OBJECT = INSTRUMENT_REFERENCE_INFO END_OBJECT = INSTRUMENT END