PDS_VERSION_ID = PDS3 LABEL_REVISION_NOTE = "Initiated JMAFI, 2004-03-01" OBJECT = INSTRUMENT INSTRUMENT_HOST_ID = P11 INSTRUMENT_ID = FGM OBJECT = INSTRUMENT_INFORMATION INSTRUMENT_NAME = "FLUXGATE MAGNETOMETER" INSTRUMENT_TYPE = "MAGNETOMETER" INSTRUMENT_DESC = " Instrument Overview =================== The High Field Fluxgate Magnetometer Experiment flown aboard the Pioneer 11 spacecraft was an extremely simple instrument used to extend the spacecraft's upper limit measurement capability by approximately an order of magnitude (from 0.14 mT to 1.00 mT) with minimum power and volume requirements. This magnetometer was designed to complement the low field measurements provided by a helium vector magnetometer and utilizes magnetic ring core sensors with bi-axial orthogonal sense coils. The instrument was a single range, triaxial fluxgate magnetometer capable of measuring fields of up to 1 mT along each orthogonal axis, with a maximum resolution of 1 microTesla. ['Instrument Overview' was adapted from ACUNA&NESS1975 p. 177.]" END_OBJECT = INSTRUMENT_INFORMATION OBJECT = INSTRUMENT_REFERENCE_INFO REFERENCE_KEY_ID = "ACUNA&NESS1975" END_OBJECT = INSTRUMENT_REFERENCE_INFO OBJECT = INSTRUMENT_REFERENCE_INFO REFERENCE_KEY_ID = "FIMMELETAL1977" END_OBJECT = INSTRUMENT_REFERENCE_INFO END_OBJECT = INSTRUMENT END