        <title>Thermo Nexus 870 FT-IR Spectrometer at RELAB</title>
        <information_model_version>1.11.0.0</information_model_version>
        <product_class>Product_Context</product_class>
        <Modification_History>
            <Modification_Detail>
                <modification_date>2020-04-09</modification_date>
                <version_id>1.0</version_id>
                <description>
                    Post-peer-review version. 
                </description>
            </Modification_Detail>
        </Modification_History>
    </Identification_Area>

    <Reference_List>
        <Internal_Reference>
            <lid_reference>urn:nasa:pds:context:facility:laboratory.relab</lid_reference>
            <reference_type>instrument_to_facility</reference_type>
        </Internal_Reference>
    </Reference_List>    
    
    <Instrument>
        <name>Thermo Nicolet Nexus 870 FTIR with Pike AutoDiff Diffuse Reflectance Accessory</name>
        <type>Spectrometer</type>
        <subtype>Near and Mid Infrared FT-IR Spectrometer</subtype>
        <description>
            Operating Spectral Range: Nominal (1.0 to 50 µm) Possible (0.8 to 200 µm)
            
            Mode  Range (µm)  Light Source  Beam Splitter   Detector
            A     1.0–4.7     Quartz	        Si-on-CaF2      TE cooled DTGS
            A*    0.9–4.5     Quartz	        Si-on-CaF2      LN2 cooled InSb
            B     1.3-28      Glowbar       Ge-on-KBr       TE cooled DTGS
            C     14.3-200    Glowbar       Solid substrate DTGS
            
            Resolution:	16, 8, 4, 2, 1 or 0.5 cm-1
            Default:	4 cm-1
            
            Geometry: 	Diffuse reflectance: Biconical, off-axis for Pike specimen holder 
            Transmission: normal incidence (collimated)
            
            Sample:	Particulate specimen or soil (reflectance):
            Measured horizontal and uncovered in purged (H2O, CO2 free) environment.
            Desired specimens amount is 50 - 500 mg.
            Specimen should be no smaller than 20 mg.
            More than 1 g is unnecessary.
            Multiple specimens can be measured in sequence (same environment)
            Chip (reflectance):
            Measured horizontal and uncovered.
            Regions to be measured should be no smaller than 1 mm.
            Maximum size is 1 cm in height and 3 cm in diameter.
            Thin section and single crystal (transmission):
            Measured vertical, attached to an appropriate aperture board.
            Regions to be measured should be no smaller than 1.5 mm.
            Surface should be polished and clean (no carbon coating).
            
            Reference Standard:	Brushed diffuse gold (reflectance)
            Blank (transmittance)
            
            Additional Attachments:	   Pike AutoDiff Diffuse Reflectance Autosampler
            Continuµm Microscope system 
            Transmission mode currently operational
            Reflectance mode (with 2-D capabilities) being tested.
            Software: 	OMNIC, Atlµs [new]
            