<?xml version="1.0" encoding="UTF-8"?>
<?xml-model href="https://pds.nasa.gov/pds4/pds/v1/PDS4_PDS_1B00.sch" schematypens="http://purl.oclc.org/dsdl/schematron"?>
<Product_Context xmlns="http://pds.nasa.gov/pds4/pds/v1"
 xmlns:pds="http://pds.nasa.gov/pds4/pds/v1"  xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance"
 xsi:schemaLocation="http://pds.nasa.gov/pds4/pds/v1 https://pds.nasa.gov/pds4/pds/v1/PDS4_PDS_1B00.xsd">
    
    <Identification_Area>
        <logical_identifier>urn:nasa:pds:context:instrument:vega2.sp1</logical_identifier>
        <version_id>1.0</version_id>
        <title>DUST IMPACT PLASMA DETECTOR for VEGA2</title>
        <information_model_version>1.11.0.0</information_model_version>
        <product_class>Product_Context</product_class>
        <Alias_List>
            <Alias><alternate_title>urn:nasa:pds:context:instrument:sp1.vega2</alternate_title></Alias>  <!-- deprecated LID -->
        </Alias_List>
        <Modification_History>
            <Modification_Detail>
                <modification_date>2021-02-24</modification_date>
                <version_id>1.0</version_id>
                <description>
                    Changed inst LIDs from
                      u:n:p:c:i:instID.scID to
                      u:n:p:c:i:scID.instID
                    And per "Guide toPDS4 Context Products" v1.7,
                    changed all lidvid_reference to lid_reference
                </description>
            </Modification_Detail>
        </Modification_History>
    </Identification_Area>

    <Reference_List>
        <Internal_Reference>
            <lid_reference>urn:nasa:pds:context:instrument_host:spacecraft.vega2</lid_reference>
            <reference_type>instrument_to_instrument_host</reference_type>
        </Internal_Reference>
        <External_Reference>
            <reference_text>
                   Reinhard, R and B. Battrick (eds), &apos;Space Missions to Halley&apos;s Comet&apos;, European
                   Space Agency ESA SP-1066, ESA Pub Div, Moordwijk, Netherlands, 1986.
            </reference_text>
            <description>reference.ESA-SP1066</description>
        </External_Reference>
    </Reference_List>
    
    <Instrument>
        <name>DUST IMPACT PLASMA DETECTOR</name>
        <type>Particle Detector</type> <!--RChen/EN was Electrode Collector-->
        <naif_instrument_id>not applicable</naif_instrument_id>

        <serial_number>not applicable</serial_number>

        <description>
 
 
    Instrument Overview
    ===================
 
      This instrument detects the electric charges
      contained in the plasma cloud generated by a solid
      particle impacting on a gold target.  The principle
      of this technique is similar to that adopted for
      the impact plasma detectors of SP-2.  The magnitude
      of the positive and negative charges is
      proportional to the mass of the particle Q=Am,
      where A= 10**3C/s for an impact velocity of 78
      km/s.  The SP-1 system includes two similar
      detectors.  Each unit is made up of a base plate
      (1) covered by a gold target (2) at zero potential,
      perpendicular to the dust flow, and an array of
      strip collectors (4).  The collectors are parallel
      to the dust flow; they are mounted on an insulator
      (3) and their edge is protected from impacts by a
      shield (5) connected electrically to the structure.
      Adjacent collectors are biased at potentials of 30
      V, with opposite polarities.  The two sets of
      electrodes in each detector detect a positive and a
      negative current pulse, which are analysed and
      recorded by the electronics unit.  The entrance to
      one detector is covered by a plastic foil, to
      obtain additional information on density and/or
      calibration factor.  The characteristic parameters
      of SP-1 are given in Table 8.
 
 
 
 
 Table 8. Dust Particle Impact Plasma Detector
 characteristics
 
 Target material               Gold (0.1 mm)
 Toil sensor area              160 cm**2, one sensor covered  with 0.6
                               and 2 micro m thick plastic foil
                               on Vegas-1 and 2, resp.
 Impact charge range          3x10**-14  -10**-8 C
 Estimated mass range         3x10**-17  -10**-11 g
 Integration time             2 s (high data rate)
                              2.5 min (low data rate)
 
 
