Name: fw_readVersion Id: 1.0.0.0
Description: The raw value from the MDIS filter wheel resolver in resolver counts. It is used by the flight software to compute mess.mdis.fw_pos. For each commanded filter number, the instrument software will try to place the filter wheel at the following positions: FILTER_NUMBER mess.mdis.fw_goal 1 17376 2 11976 3 6492 4 1108 5 61104 6 55684 7 50148 8 44760 9 39256 10 33796 11 28252 12 22852 Commanded position is reported in mess.mdis.fw_goal. There is a tolerance of 240 resolver counts around mess.mdis.fw_goal for mess.mdis.fw_pos to indicate that the filter wheel is correctly positioned.
Namespace Id: messSteward: ppiClass Name: MDISType: ASCII_​Integer
Minimum Value: 0Maximum Value: 65535Minimum Characters: NoneMaximum Characters: None
Unit of Measure Type: NoneDefault Unit Id: NoneAttribute Concept: NoneConceptual Domain: INTEGER
Status: ActiveNillable: falsePattern: None
Permissible Value(s)No Values