DESCRIPTION |
Analytical Spectral Devices LabSpec 4
https://www.malvernpanalytical.com/en/products/product-range/asd-range/labspec-range/labspec-4-hi-res-analytical-instrument
Reflectance spectra from 350 to 2500 nm (0.35 to 2.5 um) were measured at ambient temperature and pressure
relative to a calibrated Spectralon 99% diffuse reflectance standard with an Analytical Spectral Devices
LabSpec 4 Hi-Res spectrometer (ASD) at a viewing geometry of i=30°, e=0°. These spectra were corrected for the
absorption features of Spectralon. An in-house quartz-tungsten halogen light source was used to provided incident
light that was collimated to less than 1.5° divergence with a cold color temperature that provides sufficient
signal-to-noise ratio over the three sensors of the ASD instrument, with flux sufficiently low to not induce
observable emittance on very dark samples in the wavelength range under study. Five hundred scans of reflectance
spectra were collected for each sample, and 500 scans of dark current were subtracted from sample and standard
reflectance spectra.
|