DESCRIPTION |
Operating Spectral Range: Nominal (1.0 to 50 µm) Possible (0.8 to 200 µm)
Mode Range (µm) Light Source Beam Splitter Detector
A 1.0?4.7 Quartz Si-on-CaF2 TE cooled DTGS
A* 0.9?4.5 Quartz Si-on-CaF2 LN2 cooled InSb
B 1.3-28 Glowbar Ge-on-KBr TE cooled DTGS
C 14.3-200 Glowbar Solid substrate DTGS
Resolution: 16, 8, 4, 2, 1 or 0.5 cm-1
Default: 4 cm-1
Geometry: Diffuse reflectance: Biconical, off-axis for Pike specimen holder
Transmission: normal incidence (collimated)
Sample: Particulate specimen or soil (reflectance):
Measured horizontal and uncovered in purged (H2O, CO2 free) environment.
Desired specimens amount is 50 - 500 mg.
Specimen should be no smaller than 20 mg.
More than 1 g is unnecessary.
Multiple specimens can be measured in sequence (same environment)
Chip (reflectance):
Measured horizontal and uncovered.
Regions to be measured should be no smaller than 1 mm.
Maximum size is 1 cm in height and 3 cm in diameter.
Thin section and single crystal (transmission):
Measured vertical, attached to an appropriate aperture board.
Regions to be measured should be no smaller than 1.5 mm.
Surface should be polished and clean (no carbon coating).
Reference Standard: Brushed diffuse gold (reflectance)
Blank (transmittance)
Additional Attachments: Pike AutoDiff Diffuse Reflectance Autosampler
Continuµm Microscope system
Transmission mode currently operational
Reflectance mode (with 2-D capabilities) being tested.
Software: OMNIC, Atlµs [new]
|