Instrument Information
IDENTIFIER urn:nasa:pds:context:instrument:sbu_cpex.raman::1.0
NAME RAMAN SPECTROMETER
TYPE SPECTROMETER
DESCRIPTION
Micro-Raman spectra for this project were obtained by the Center for Planetary Exploration at 
            Stony Brook University using a a WiTec alpha300R confocal imaging system equipped with 532 nm 
            neodymium: yttrium/aluminum/garnet (Nd:YAG) laser with 2.24 mW nominal power at the sample 
            surface, and a 50X (NA = 0.8) objective (spot size of 763 nm). Each analysis consisted of 240 
            one-second integrations that were averaged. For Raman and micro-FTIR spot analyses, multiple 
            locations were analyzed per sample.
MODEL IDENTIFIER
NAIF INSTRUMENT IDENTIFIER
SERIAL NUMBER
REFERENCES Jaret, S.J., Johnson, J.R., Sims, M., DiFrancesco, N., and Glotch, T.D., Microscopic and petrographic comparison of experimentally shocked albite, andesine, and bytownite, Journal of Geophysical Research, 123, 1701-1722, 2018, doi:10.1029/2018JE005523.